Update log

Change log

[2023.03.14] Samplecase is installed on AVNEO400
- Before : manual injection --> After : autosampler

[2023.03.14] Samplecase-plus is installed on AS400
- Before : 24 holder --> After : 60 holder

[2022.09.27] ATMA firmware is updated on AV500

[2022.09.27] ATMA firmware is updated on AVHD400 

[2022.08.30] Probe is upgrade on AVHD400
- Before : BBFO probe --> After : iProbe 
- better : Signal-to-noise ratio is enhanced about 1.5 ~ 2 times
- worse : 0.11% linewidth is broader (Shim : Boss I) 
- 1H, 13C, 19F, 31P, 15N based experiments are possilbe (2022.08.30)
- 7Li, 11B, 29Si is available (2022.08.31)
- High temperature correction is updated (2022.09.01) 
- Low temperature correction will be updated 
- other heteronuclei will be updated on demand

[2022.07.14] Topspin is upgraded on AVHD400 
- Topspin 3.6.5 on CentOS 7 

[2022.07.12] Topspin is upgraded on AS400 
- Topspin 3.6.5 on CentOS 7 

[2022.07.12] Topspin is upgraded on AVNEO400 
- Topspin 4.1.4 on CentOS 7 
# atma firmware is updated 

[2022.07.04] Topspin is upgraded on AV500
- Topspin 4.1.4 on CentOS 7 
# Bugfix : trouble on 19F atma is fixed 
# atma firmware is updated 

[2021.12.13] Topspin is upgraded on AV500
- Topspin 4.1.3 on CentOS 7 
# Bugfix : trouble on 19F atma is fixed 
# BSMS(ELCB) firmware is updated

[2021.11.11] Minor change on Workstation of AV500
- Additional network card(realtek) is attached 
# After recent CentOS7 update, two NIC (Intel) are not working together due to different subnet address (maybe driver problem) 
# No problem is found on Window10 and Ubuntu, so lan driver or lan configuration of centOS7 are suspected. 

[2021.11.09] AVNEO400 is upgraded (former AV400)
- Probe : iProbe
- Console : AVNEO
- Magent : ASCEND 400
- Software : Topspin4 on CentOS 7

[2021.11.06] Workstation of AVHD400 is upgraded
- Workstation : HP Z4 G4 (former HP Z420)
- Topspin 3.6.3 on CentOS 7 

[2021.10.28] AV400 is decommissioned for upgrade
- Probe : move to AS400 
- Magnet : Quenched (Shim coil is broken)

[2021.10.28] AS400 is under repair service
- Dual probe(1H/13C) is installed 
[2021.10.21] AS400 is updated
- Topspin is upgraded to 3.6.3 for iProbe compatibility
- iProbe is installed until AV400 is upgraded 

[2021.10.01] Topspin is upgraded to 3.6.3 (AVHD400)
- for iProbe compatibility 
- 'apbk' command is available for automatic phase & baseline correction together
- Codemeter license 

[2021.09.01] AV300 is decommissioned 
- Console : CCU is not working
- Probe : Quartz is broken
- Magnet : Quenched

[2021.06.02] atma is updated for 19F tune (AV500)

[2021.05.25] Topshim trouble is fixed (AV500)

[2021.05.13] new experiments are imported to AV500
- 19F_IGD_noBG.kaist : 19F experiment with inverse-gated decoupliing (Background signal is not shown)
- 19F_noCPD_noBG.kaist : 19F experiment (Background signal is not shown)

[2021.04.29] new experiments are imported to AV500
- Pureshift.kaist : 1H spectrum without multiplet 
# Full range 1H homo-decoupling 

[2021.04.21] AV500 is updated
- Software : Topspin 4.1.1 (before 4.0.9)
- OS       : CentOS 7 (Not changed) 

[2021.02.19] parameter of 2D experiments(NUS) are updated
- set NusPoint more than 48 (more than 64 recommended) 
- set SW (1H) as 13 ~ 14 ppm 
- set O1P (1H) as 6ppm 

[2021.02.19] Full automation test on AV500 until May
- ICON-NMR is running between 17:00(Fri) ~ 08:30(Mon)

[2021.02.16] Web page is updated 
- set 'Reservation Start time' at 9:00 (before 08:00) 
- new 'long experiment' reservation rule at lunch/dinner time
- set 'Weekend timetable' differently to 'Weekday timetable'
- user can view reservation history

[2021.02.01] backup probe is installed on AV500 
- iProbe : 19F atma trouble on 1H-19F HOESY --> under service
- BBFO probe is installed (backup)

[2020.11.01] Mestrenova license server is updated
- IP address is changed
- updated Campus license files are uploaded on Data server 
# KAIST_Campus_License_2020.zip

[2020.10.07] AVHD400 is updated
- Software : Topspin 3.5 (before 3.1)
- OS       : CentOS 7 (before 5)
- Workstation : Z420 (before Z400)
- Firmware